Instrument functions
R&S
®
NGU
68
User Manual 1179.2531.02 ─ 03
Figure 6-34: Safety limits dialog
2. Activate the "Enabled" menu item.
The R&S
NGU limits the set voltage and current level and displays the "Safety Lim-
its" icon on the selected channel status bar information.
3. Set the required minimum and maximum value for voltage and current level.
The R&S
NGU displays an on-screen keypad to set the value.
4. Confirm value with the unit key.
6.8.5
Oscillation detection
Depending on load impedance and mode setting, the control-loop may start to oscil-
late. To protect the load, an oscillation detector will trigger and deactivates the output. If
oscillation occurs, set to slow transient response or select "Capacitance Mode" accord-
ing to the load impedance for improved stability.
Chapter 6.4, "Capacitance mode"
Chapter 6.2.2, "Fast transient
6.9
Trigger / Digital I/O
Digital I/O pins voltage rating
Do not exceed the maximum voltage rating of the Digital I/O pins when supplying vol-
tages to the pins.
The specified voltages are 0 V to 24 V for all output pins and 0 V to 15 V for all input
pins.
For more information, see the instrument datasheet.
Trigger / Digital I/O