Remote control commands
R&S
®
NGU
165
User Manual 1179.2531.02 ─ 03
Example:
ARB:PRI:MODE VPM
Set priority mode to VPM.
ARB:DATA 10,1,-1,0.5,0
Defines one arbitrary point with: voltage1 = 10 V and positive
current1 = 1 A, negative current1 = -1A, time1 = 500 ms and
interpolation mode1 = 0 (disabled).
ARB:DATA?
-> 10.000, 1.000, -1.000, 0.5, 0
Returns defined arbitrary points for model R&S NGU201 in VPM
priority mode.
Returns defined arbitrary points for the previous selected chan-
nel.
Example:
Example "Configuring an arbitrary sequence"
ARBitrary:FNAMe
<filename>[,<location>]
ARBitrary:FNAMe?
[<location>]
Sets or queries the file name and storage location for the arbitrary function.
Parameters for setting and query:
<filename>
Filename of the arbitrary function.
<location>
INT | EXT | DEF
INT
Internal memory
EXT
USB stick
DEF
Internal memory
Example:
ARB:FNAM "01.CSV"
ARB:FNAM? INT
-> "01.CSV"
ARBitrary:LOAD
Loads an arbitrary table from a file (filename specified with
ARB:FNAM
)
Example:
ARB:DATA 10,1,0.5,0
ARB:REP 10
ARB:FNAM "ARB03.CSV“,INT
ARB:SAVE
ARB:LOAD
Loads an arbitrary data from filename
ARB03.CSV
.
Usage:
Event
Advanced operating commands