92
5975 Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Connect the Micro-Ion Vacuum Gauge
The Micro-Ion vacuum gauge is optional for EI operation. It is
required
for
operating the MSD in CI mode.
Materials needed
•
Micro-Ion vacuum gauge (G3397A)
•
Power cord
•
Micro-Ion vacuum gauge cable (G3170-60805)
•
CI flow module cable-Smart Card to CI flow (G3170-60808)
Procedure
1
Use the provided KF-16 clamp to install the Micro-Ion vacuum gauge to the
rear of the analyzer chamber.
2
Connect the gauge cables as shown in
.
3
Pump down the MSD.
Summary of Contents for 5975C TAD VL MSD