General Troubleshooting
2
5975 Series MSD Troubleshooting and Maintenance Manual
49
High foreline pressure
•
Excessive carrier gas flow (typically > 5 mL/min)
•
Excessive solvent volume injected
•
Large vacuum leak
•
Severely degraded foreline pump oil (standard foreline pump)
•
Collapsed or kinked foreline hose
•
Foreline pump is not working correctly
•
Foreline gauge is not working correctly (diffusion pump MSD only)
Internal MS communication fault
•
MSD electronics are not working correctly
Lens supply fault
•
Electrical short in the analyzer
•
MSD electronics are not working correctly
Log amplifier ADC error
•
MSD electronics are not working correctly
No peaks found
•
Emission current was set to 0
•
Electron multiplier voltage is too low
•
Amu gain or offset is too high
•
Poor mass axis calibration
•
Amu gain or offset is too high
•
Calibration vial(s) empty or almost empty
•
Excessive pressure in the analyzer chamber
•
Air leak
•
Electron multiplier voltage is too low
•
Signal cable is not connected
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...