5975 Series MSD Troubleshooting and Maintenance Manual
8
Electronics
Side Board
The side board is mounted on the side plate. It performs these functions:
•
Provides the 1
MHz reference clock for the RF amplifier.
•
Generates the RF component of the voltage applied to the quadrupole mass
filter according to a signal from the main board. The amplitude of this
voltage is proportional to the mass selected.
•
Generates the DC component of the voltage applied to the quadrupole mass
filter. The magnitude of this voltage is proportional to the RF voltage.
•
Passes voltages generated on the main board and the detector focus voltage
from the HED power supply to elements in the ion source and the detector.
•
Generates and adjusts filament emission current and electron energy as
controlled by the main board.
•
Switches the filament power from one filament to the other.
•
Monitors for RF faults and shuts down the analyzer if one is detected.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...