![Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual Download Page 183](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605183.webp)
Vacuum System
6
5975 Series MSD Troubleshooting and Maintenance Manual
183
Compression seals
A compression fitting consists of a threaded fitting on the analyzer chamber
and a threaded collar with a ferrule and O-ring. A cylindrical part fits inside
the collar. Tightening the collar presses the ferrule, compressing the O-ring
around the part. The calibration vials use compression seals.
High voltage feedthrough seal
The high voltage (HED) feedthrough seal is an O-ring that is compressed
against the side plate by a threaded collar.
Figure 38
Vacuum seals
Side plate O-ring seal
KF seal with internal
centering ring
KF seal with external
centering ring
Compression seal
(clamp not shown)