Analyzer
7
5975 Series MSD Troubleshooting and Maintenance Manual
207
Entrance lens
The entrance lens (
) is at the entrance to the quadrupole mass filter.
This lens minimizes the fringing fields of the quadrupole which discriminate
against high-mass ions. There is a per4.4
volt voltage added to the
entrance lens. The total voltage applied to the entrance lens is the sum of the
entrance lens offset and entrance lens gain and the +4.4
volt
permanent
offset.
Entrance lens voltage = +4.4
VDC
+
offset
+
(gain
×
mass)
Entrance lens offset
The entrance lens offset (EntOff) controls the fixed voltage applied to the
entrance lens. It can be varied from 0 to –64
VDC
(–20
V
is
typical).
Increasing
the entrance lens offset generally increases the abundance of ions at low
masses without substantially decreasing the abundance of high mass ions.
Entrance lens gain
Entrance lens gain (EntLens) controls the variable voltage applied to the
entrance lens. It determines how many volts are applied for each
m/z
. It can be
varied from 0 to –128
mV/(
m/z
). A typical range is 0 to –40
mV/amu.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...