General Maintenance
To Lubricate the Vent Valve O-Ring
Materials needed
•
Cloths, clean (05980-60051)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Grease, Apiezon L, high vacuum (6040-0289)
•
O-ring, vent valve (0905-1217). Replace if the old O-ring is worn or damaged
The vent valve O-ring needs a thin coat of lubrication to ensure a good vacuum
seal and smooth operation. If the vent valve O-ring does not turn smoothly or
does not seal correctly, lubricate it using this procedure.
Procedure
1
Vent the MSD.
2
Completely remove the vent valve knob (
).
3
Inspect the O-ring. If the O-ring appears damaged, replace it.
4
Use a clean, lint-free cloth or glove to spread a
thin
coat of high vacuum
grease on the exposed surface of the O-ring.
5
Use a clean, lint-free cloth or glove to wipe away excess grease. If the
O-ring looks shiny, there is too much grease on it
C A U T I O N
Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be
lubricated. Lubricating other seals can interfere with their correct function.
C A U T I O N
Excess grease can trap air and dirt. Grease on surfaces of the O-ring other than the
exposed surface can trap air, resulting in air spikes during operation.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...