![Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual Download Page 220](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605220.webp)
220
5975 Series MSD Troubleshooting and Maintenance Manual
8
Electronics
•
Supplies and controls the power for the ion source and quadrupole (mass
filter) heaters.
•
Provides 24
VDC power for the cooling fans.
Signal amplifier board
The signal amplifier board amplifies the output of the detector. It produces an
output voltage of 0 to 10
volts DC, proportional to the logarithm of the input
current of 3
picoamps to 50
microamps.
An analog-to-digital converter converts the amplifier output voltage to digital
information. The LAN/MSD control card converts the data into abundance
counts proportional to the detector signal current.
AC board
The AC board is mounted on the opposite side of the electronics panel from
the LAN/MSD control card. The AC board is also sometimes called the power
distribution/vacuum control board. It performs these functions:
•
Provides input voltage transparency for the MSD.
•
Distributes AC line power to the AC/DC power supply, the foreline pump,
and the turbo pump controller.
•
Turns the calibration valve on or off as directed by the main board.
•
Provides the voltage for the calibration valve.
•
Provides a logic interface to turbo controller.
•
Controls the diffusion pump:
•
Controls the foreline gauge.
•
Turns on the diffusion pump once the foreline pressure is low enough, as
directed by the main board.
•
Regulates the AC power to the diffusion pump heater.
•
Turns off the diffusion pump if the foreline pressure is too high or if the
diffusion pump is too hot.
•
Passes the foreline pressure signal from the foreline gauge or turbo pump
speed and other vacuum status information to the main board.
•
Turns off the foreline pump in case of a problem with pumpdown.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...