5975 Series MSD Troubleshooting and Maintenance Manual
3
CI Troubleshooting
Cannot complete Autotune
Wrong or corrupted tune file
Check the tune parameters.
The
m/z
28/27 ion ratio (for methane) is incorrect. The correct ratio should be
between 1.5 and 5.0.
If the ion ratio is incorrect, adjust it. See the
Operation Manual.
The CI ion source is dirty.
Clean the ion source. See
164 for more
information.
Too much air or water in the system
See
60. After eliminating these problems, clean the ion
source.