60
5975 Series MSD Troubleshooting and Maintenance Manual
3
CI Troubleshooting
Air Leaks
How do I know if I have an air leak?
Large
air leaks can be detected by vacuum symptoms: loud gurgling noise
from the foreline pump, inability of the turbo pump to reach 95% speed, or, in
the case of smaller leaks, high pressure readings on the high vacuum gauge
controller.
The mass flow controller is calibrated for methane and the high vacuum gauge
controller is calibrated for nitrogen, so measurements are not accurate in
absolute terms:
Familiarize yourself with the measurements on
your
system under operating
conditions. Watch for
changes
that may indicate a vacuum or gas flow
problem.
Always
look for small air leaks when setting up methane flow. Run the
methane
pretune, starting with a good PCI tune file (
of
m/z
19 (protonated water) should be less than 50% of
m/z
17 for acceptable
PCI performance. For NCI, the abundance of
m/z
19 (protonated water)
should be less than 25% that of
m/z
17. If the MSD was just pumped down, look
for the abundance of
m/z
19 to be decreasing.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...