![Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual Download Page 144](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605144.webp)
144
5975 Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To reinstall a filament
Materials needed
•
Filament assembly (G2590-60053)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
Procedure
1
Install the new filament (
2
Remove the screw holding the filament to the ion source body.
3
Slide off the filament assembly.
4
Orient the assembly so that the filament is next to the ion source body.
5
Replace the screw to the ion source body.
6
After installing the filament, verify that it is not grounded to source body.
7
Reinstall the ion source. See the
Operation Manual
.
8
Close the analyzer chamber.
9
Pump down the MSD.
10
Autotune the MSD.
11
In the Edit Parameters dialog box (Instrument/Edit MS Tune Parameters),
select the other filament.
12
Autotune the MSD again.
13
Select and use the filament that gives the best results.
If you decide to use the first filament, run Autotune again to make sure the
tune parameters are compatible with the filament.
14
Select
Save Tune Parameters
from the File menu.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...