![Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual Download Page 24](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605024.webp)
24
5975 Series MSD Troubleshooting and Maintenance Manual
1
Introduction
Operating precautions
•
Turn off the hydrogen at its source every time you shut down the GC or
MSD.
•
Turn off the hydrogen at its source every time you vent the MSD (do not
heat the capillary column without carrier gas flow).
•
Turn off the hydrogen at its source every time shutoff valves in an MSD are
closed (do not heat the capillary column without carrier gas flow).
•
Turn off the hydrogen at its source if a power failure occurs.
•
If a power failure occurs while the GC/MSD system is unattended, even if
the system has restarted by itself:
1
Immediately turn off the hydrogen at its source.
2
Turn off the GC.
3
Turn off the MSD and allow it to cool for 1 hour.
4
Eliminate
all
potential sources of ignition in the room.
5
Open the vacuum manifold of the MSD to atmosphere.
6
Wait at least 10 minutes to allow any hydrogen to dissipate.
7
Start up the GC and MSD as normal.
When using hydrogen gas
check the system for leaks to prevent possible fire
and explosion hazards based on local Environmental Health and Safety (EHS)
requirements. Always check for leaks after changing a tank or servicing the
gas lines. Always make sure the vent line is vented into a fume hood.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...