![Agilent Technologies 5975C TAD VL MSD Troubleshooting And Maintenance Manual Download Page 159](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605159.webp)
General Maintenance
4
5975 Series MSD Troubleshooting and Maintenance Manual
159
To adjust the quad frequency
Materials needed
•
Screwdriver, flat-blade, large (8730-0002)
Procedure
1
Make sure the MSD is at thermal equilibrium. It takes at least 2
hours
after
all heated zones have reached their setpoints for the MSD to reach
thermal equilibrium.
2
Open the analyzer cover.
3
Make sure the RF cover on the side board is secure and no screws are
missing. A loose RF cover or missing screw can
significantly
affect coil
adjustment.
4
In the Tune and Vacuum Control view, select
Optimize Quadrupole Frequency
from the Execute menu
.
5
Enter an
m/z
value of 100.
6
Slowly turn the quad frequency adjustment screws to minimize the voltage
displayed (
Turn the adjustment screws alternately. Turn each screw only a little bit at
a time. Keep the screws at
equal
extension. The minimum voltage is
typically between 50 and 70
mV.
7
When the voltage is minimized, click
Stop
.
WA R N I N G
Do not remove any other covers. Dangerous voltages are present under these
covers.
C A U T I O N
Do not use a coin to adjust the screws. If you drop it, it could fall into the electronics
fan and cause significant damage.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...