General Maintenance
4
5975 Series MSD Troubleshooting and Maintenance Manual
101
To Check the Diffusion Pump Fluid
Materials needed
•
Screwdriver, Torx T-15 (8710-1622)
Procedure
1
Remove the analyzer window cover.
2
Vent the MSD.
3
Remove the side cover.
4
Check the diffusion pump fluid level. See
.
The diffusion pump fluid level can be seen through the window below the
fan at the front of the MSD. The diffusion pump fluid level should be
between the top and bottom of one of the
FULL
ranges. There are two sets of
marks. Use the
HOT
marks if the diffusion pump is on and is at its normal
operating temperature. Use the
COLD
marks if the pump is off and has had
time to cool. If the fluid level is below the bottom of the appropriate range,
replace the diffusion pump fluid. Do not just add fluid.
The pump fluid should be clear or almost clear. Dark or cloudy pump fluid
indicates an air leak or excessive heat. If the pump fluid appears dark or
cloudy, replace it. Then, check for an air leak.
The diffusion pump fluid should be replaced at least once a year, or more
often if the pump fluid level is low or if the fluid is dark or cloudy.
WA R N I N G
Do not remove any other covers. Removing other covers may expose hazardous
voltages.
WA R N I N G
Keep your hair away from the cooling fan if the MSD is turned on.
Summary of Contents for 5975C TAD VL MSD
Page 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Page 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Page 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Page 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...