5975 Series MSD Troubleshooting and Maintenance Manual
2
General Troubleshooting
•
Calibration valve is not working correctly
•
Bad signal cable connection
•
Filament has failed or is not connected correctly
•
Bad ion source wiring connection
•
Bad detector wiring connection
•
Failed electron multiplier horn
Peaks are tailing
•
Active sites in the sample path
•
Injection is too large
•
Incorrect GC inlet temperature
•
Insufficient column flow
•
GC/MSD interface temperature is too low
•
Ion source temperature is too low
Peaks are fronting
•
Column film thickness mismatched with analyte concentration (column
overload)
•
Initial oven temperature is too low
•
Active sites in the sample path
•
Injection is too large
•
GC inlet pressure too high
•
Insufficient column flow
Peaks have flat tops
•
Insufficient solvent delay
•
Incorrect scale on the display
•
Injection is too large
•
Electron multiplier voltage is too high