CAUTION:
In case of functional defects or other deviations from normal
operation behavior the unit has to be switched off immediately
and the service to be informed. The equipment must only be put
back into operation when the fault has been repaired.
CAUTION:
The use of spare parts from third party suppliers can affect the
safety of the equipment. If components fail, use only original
spare parts.
Table 5: Maintenance
Lifetime
Expected lifetime for the X-ray system
10 years
Periodic maintenance
In order to ensure good and safe functioning conditions
of the equipment, preventive maintenance must be car-
ried out. If you have not signed a maintenance contract,
please contact the manufacturer’s service department or
its agent.
Every 12 months
Maintenance by the user
Clean any residual contrast mediums from the tabletop. Daily
Do a visual check of the equipment movement, the dis-
plays and the luminous signals.
Daily
Make sure that the necessary accessories for immobiliz-
ing the patient are properly fixed to the equipment, and
particularly the handgrips.
Daily
Check the correct functioning of the ceiling suspension:
• Verify that no error messages appear on the display.
• Push the movement stop button and perform a
movement of the system.
• Check if movements can be performed easily and
without impedance.
• Check the proper functioning of the collimator unit.
Daily
Check that the movement stop buttons function proper-
ly.
Every 6 months
Check the dimensions of the luminous field delimited by
the collimator by following the procedure included in
the technical manuals.
Every 6 months
78
| DR 800 | Introduction
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...