1
2
4
3
3
1
3
3
2
4
1.
Patient orientation
2.
Mounting unit
3.
Tabeltop guides
4.
FLFS overlay
Figure 126: FLFS overlay on radiographic table (head left, head right)
3.
Use the positioning console to move the X-ray system to the automatic
position.
The X-Ray system moves to the default position for preparing the DR Full
Leg Full Spine examination.
If the radiographic table is tilted to vertical position, assure that the FLFS
overlay remains flat against the tabletop.
CAUTION:
After doing vertical FLFS exposures the grid can swing away
from the table and disturb the patient. Set the inclination of
the table to 89 instead of 90 degrees.
162
| DR 800 | Full Leg Full Spine
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...