Figure 94: Digital tomosynthesis controls
3.
Move the X-ray system to the right position.
a) Check if a correct automatic position is selected.
Figure 95: Positioning controls on the software console
b) Move the X-ray system to the selected automatic position.
The actual and target position parameters are displayed on the
software console. When the target position is reached, the movement
stops.
c) Adjust the position using the position controls.
4.
Position the patient.
The patient position can be verified using the collimator camera.
WARNING:
Warn the patient that the X-ray tube will perform a sweeping
movement during the examination. Give instructions to
avoid that the patient looses balance and to avoid injuries in
patient hands or fingers.
5.
On the collimator, switch on the light localizer. Apply collimation.
6.
Acquire a static image.
If a reference image is required, acquire a static image. The images of the
acquisition sequence should not be used to replace a static image.
Press and hold down the exposure button or the radiography pedal to
make an exposure to acquire a static image.
The image is stored and displayed as a thumbnail in the lower half of the
Image Overview
pane.
Figure 96: Thumbnail of a static image
If required, multiple static images can be made.
Depending on the configuration, the acquisition of static images during a
DR workflow for digtial tomosynthesis may not be possible.
7.
In the digital tomosynthesis screen of the software console, click the
button to start the digital tomosynthesis workflow.
126
| DR 800 | Basic Workflow
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...