One point, two point and three point working modes
You can select following radiographic working modes according to the
parameters to be controlled and the degree of automation:
Table 28: Radiographic working modes
mAs
mA,ms
One point mode, by selecting kV. The
exposure is controlled by AEC.
mAs
mA,ms
Two point mode, by selecting kV and
mAs. AEC is disabled.
mAs
mA,ms
Three point mode, by selecting kV,
mA and exposure time independent-
ly. AEC is disabled.
To switch to one point mode, activate one or more AEC fields.
Depending on the radiographic working mode, some of the generator controls
will be disabled.
208
| DR 800 | Software Console
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...