Compressor cannot be released
De-
tails
A patient is lying on the table and the compressor is lowered, but
cannot be released using the position controls.
Cause Power failure.
System failure that prevents the remote control of the compressor.
Brief
Solu-
tion
1.
Fetch the wrench.
The wrench is attached to the cable support turret.
Figure 176: Cable support turret with wrench attached to it
2.
Turn the bolt on the rear side of the compressor carriage coun-
terclockwise (seen from the rear).
The compressor moves upward as the bolt is turned.
Figure 177: Rear view of the compressor carriage
264
| DR 800 | Problem solving
0392C EN 20210309 1049