WARNING:
To avoid risk of electric shock, this equipment must only be
connected to a grounded electrical network.
Ceiling suspension
The system is not supplied with any type of ON/OFF switch control or related
circuitry. It is required of the installer to complete this external circuit and
instruct the operators on its use and operation once installation is complete.
The stand should be incorporated into the examination room general
emergency shutdown system.
The equipment is factory set for using single-phase mains rated at 230 VAC
50/60 Hz with a maximum power consumption of 300 VA. The system is also
prepared to accept a load from the manual or automatic collimation system
with lamp ratings up to 150 W maximum.
The stand must be connected to a stable power source to obtain the
performances for which it is designed.
Table 62: Power input requirements
T1-VOLTAGE SUPPLY
230 VAC
F1-FUSE TYPE
6.3 x 32 mm
F1-FUSE RATING
T 1.6 A
VA
300
Located on the main electrical plate is power strip terminal [ST2]. This is
provided for aftermarket connection to various collimation systems. The
circuit feeds 24 VAC and is able to power various collimation systems,
including lamps, rated up to 150 watts. Verify that the collimation system or
accessory does not overdraw the current rating on the circuit.
Table 63: Auxiliary power output
T1-VOLTAGE SUPPLY
24 VAC
F1-FUSE TYPE
6.3 x 32 mm
F1-FUSE RATING
T 6.3 A
VA
150
The manufacturer reserves the right to make changes to T1, fuse type, and
ratings (for optimization, enhancements, and in accordance to government
regulations regarding components used in medical devices). For the latest
information regarding voltages and ratings, always refer to the electrical
schematics that come with the equipment.
308
| DR 800 | Product Information
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...