Voltage dips,
short interrup-
tions and voltage
variations on
power supply iput
lines in accord-
ance with IEC
61000-4-11
0% U
n
for 0.5 cy-
cle
0% U
n
for 1 cycle
70% U
n
for 25
cycles
0% U
n
for 5 s
IEC 60601-1-2
Test level
Mains power quality
should be that of a
typical commercial
or hospital environ-
ment. If the user of
the system requires
continued operation
during power mains
interruptions, it is
recommended that
the system be pow-
ered from an unin-
terruptible power
supply or a battery.
Magnetic field at
the supply fre-
quency (50/60
Hz) in accordance
with IEC
61000-4-8
30 A/m
IEC 60601-1-2
Test level
Magnetic field at the
network frequency
should correspond
to the typical values
as they are in a com-
mercial and clinical
environment.
• REMARK : U
n
is the alternating current in the network before the appli-
cation of the test level.
Immunity to RF wireless communication equipment
ISM Band
(MHz)
Service
Distance
(m)
Immunity
test level
(V/m)
300-390
TETRA 400
0.3
27
430-470
GMRS 460; FRS 460
0.3
28
704-787
LTE Band 13, 17
0.3
9
800-960
GSM 800/900; TETRA 800, IDEN
820; COMA 850; LTE Band 5
0.3
28
1700-1990 GSM 1800; COMA 1900; GSM 1900;
DECT; LTE Band 1, 3, 4, 25; UMTS
0.3
28
2400-2570 Bluetooth; WLAN; 802.11 b/g/n;
RFID 2450; LTE Band 7
0.3
28
5100-5800 WLAN 802.11 a/n
0.3
9
314
| DR 800 | Product Information
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...