DR workflow with fluoroscopy for positioning
This workflow is available only on DR systems that support dynamic imaging.
Fluoroscopy can be used as a guide for positioning the patient before
performing the planned exposure.
To use fluoroscopy for positioning:
1.
Add a fluo group to the
Image Overview
pane.
If a fluo group has already been added based on data from the RIS, this
step can be skipped.
a) In the
Examination
window, click
Add Image
.
The
Add Image
window appears.
Figure 62: Add Image
b) Specify the exam group and exam type by clicking on the buttons.
c) Select an exam type that is configured as a fluo group and click
OK
.
The fluo group thumbnail is added to the
Image Overview
pane.
A fluo group thumbnail is indicated with an icon in the top right corner of
the thumbnail.
Figure 63: Thumbnail for a fluo group
2.
Select the thumbnail for the fluo group in the
Image Overview
pane of
the
Acquisition
window.
The selected DR detector is activated. The default X-ray exposure
parameters and X-ray system position for the selected examination are
sent to the modality. The software console displays these settings in the
examination overview.
DR 800 | Basic Workflow |
111
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...