The workflow for acquiring a DSA sequence is explained in the next steps.
The other workflows are explained in other sections in this manual.
7.
Select the DSA mode in the software console.
Figure 109: DSA mode
WARNING:
The DSA image processing relies on absence of movement.
Do not change the position of the table, X-ray tube or
collimator during the DSA examination.
8.
Press and hold down the radiography pedal or the exposure button.
A first set of frames is used to compose the mask image. The subsequent
frames are displayed with the mask image subtracted. The syringe icon
indicates that the mask image has been created.
DR 800 | Basic Workflow |
133
0392C EN 20210309 1049