Radiographic Parameters
You can set up following radiographic parameters:
•
kV
: shows the radiographic kV value (X-ray tube voltage) selected for the
exposure.
•
mAs
: shows the radiographic mAs value selected for the exposure.
•
mA
: shows the radiographic mA value (current) selected for the exposure.
•
ms
: shows the time value (in milliseconds) selected for the exposure.
•
Max ms
: shows the integration time of the DR detector. When operating
the DR detector, the calculated exposure time (ms) or manual overrides
can never exceed the integration time (detector ms) of the DR detector.
•
Max mAs
: shows the maximum allowed mAs value for exposures using
AEC. The highest allowed setting for max mAs depends on the mA setting
and the detector ms setting. Only available in AEC mode.
When an exposure is made, the actual exposure parameters are displayed in a
message.
When using AEC, the exposure is terminated by the detector ms or max mAs
settings or by the AEC dose failure safety device, even if the target dose is not
reached.
Related Links
on page 239
Automatic Exposure Control (AEC)
on page 212
on page 218
Extra parameters in rapid sequence mode
Following parameters are only available in rapid sequence mode:
•
FPS
: frames per second
•
Binning
: when capturing dynamic images the amount of data to be
transferred can be reduced by lowering the resolution of the captured
images. This is achieved by combining the readouts of adjacent pixels into
a single value and this process is called binning.
•
Max RS
shows the maximum time that is available to perform rapid
sequence exposures, based upon the load of the X-ray tube and the heat
unit status.
Related Links
210
| DR 800 | Software Console
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...