Prepare the Full Leg Full Spine configuration
Position the X-Ray system and the FLFS overlay:
1.
Place the tabletop in the horizontal position.
2.
Position the FLFS overlay on the radiographic table.
a) Insert the mounting unit on the tabeltop guide.
If the patient is positioned with the head on the left side, the mounting
unit must be inserted on the rear guide. If the patient is positioned
with the head on the right side, the mounting unit must be inserted on
the front guide.
b) Fasten the mounting unit by turning both the handle and the smaller
knob thoroughly clockwise.
1
2
4
3
3
1
3
3
2
4
Figure 124: Fastening the mounting unit (head left)
c) Put the overlay on the tabletop and slide it into the mounting unit until
it automatically fixes.
1
2
4
3
3
1
3
3
2
4
Figure 125: Mounting the overlay (head left)
NX is configured for a specific patient orientation, either head left
(default) or head right.
In longitudinal direction the position of the overlay on the tabletop can be
chosen depending on the reqion of interest of the planned examination.
The overlay should not protrude from the sides of the tabletop.
DR 800 | Full Leg Full Spine |
161
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...