Radiation status
Table 25: Radiation status
Icon Description
This icon indicates that there is no specific radiation status to be re-
ported. By touching the icon, the exposure controls of the system are
disabled, regardless of the ready for exposure status.
The exposure controls of the system are disabled. No exposure can
be made. When this status is active, the X-ray modality status frame
is highlighted with a hatch pattern. Touch the icon to release the ex-
posure controls.
The exposure button is pressed halfway and the X-ray tube is pre-
pared.
The exposure button is pressed completely and the X-ray exposure is
made. The indicator on the console will light up.
The examination room door is open.
Press the exposure button halfway (“Prep” position) to prepare the X-ray tube
for exposure. The indicator will light up when the X-ray tube is prepared and
there are no interlock failures or system faults.
After pressing the exposure button, the following functions are activated:
• Anode rotation.
• Filament current switches from stand-by to the selected mA.
200
| DR 800 | Software Console
0392C EN 20210309 1049