Button Parameter
Indication that automatic stitching has been performed based on
the alignment of the anatomical information in the image.
Indication that patient movement is detected.
Indication that automatic stitching has been performed based on
the stitching grid.
Indication that manual corrections have been applied to the align-
ment of the partial images.
CAUTION:
Patient movement can cause inaccurate alignment of the partial
images. Patient movement is not always detected by the system.
The user is responsible for observation of the movement of the
patient during the examination.
To perform quality control:
1.
If required, adjust the stitching.
2.
Click
Accept
.
Depending on the configuration settings, the stitching parameters are
added to the image as a text annotation.
The text annotation contains following information:
Table 16: Annotations
V Vertical correction, in respect to the stitching grid, for the alignment
of the anatomical information in the image.
H Horizontal correction, in respect to the stitching grid, for the align-
ment of the anatomical information in the image.
DR 800 | Full Leg Full Spine |
171
0392C EN 20210309 1049