DR Detector must be Recalibrated
Details
A message is displayed on NX indicating that the DR De-
tector must be recalibrated.
Cause
A DR Detector must be recalibrated at regular interval.
Brief Solution
Follow the instructions in the DR System Key User Man-
ual to calibrate the DR Detector:
• DX-D DR Detector Calibration Key User Manual,
document 0134
DR 800 | Problem solving |
265
0392C EN 20210309 1049