10.
Source image distance (SID). This distance is received together with the X-
Ray generator parameters.
Figure 147: Making measurements on DR Full Leg Full Spine images
The respective distances between the plane in which the measurement is
made, the stitching grid, the DR Detector and the X-ray tube are used to
calculate the Estimated Radiographic Magnification Factor to calibrate the
distance measurement on NX.
186
| DR 800 | Full Leg Full Spine
0392C EN 20210309 1049