Mounting the lateral cassette/detector holder
WARNING:
Both the mounting and the use of the lateral cassette/detector
holder are allowed only with the table in the horizontal
position. In no case is the operator allowed to use this accessory
with the table in vertical position.
A DR detector can be mounted in landscape or portrait position, with or
without a handle unit.
1.
Place the table in the horizontal position.
2.
Mount the detector holder on the table.
a) Loosen the knob on the mount of the lateral detector holder.
b) Insert the detector holder on the tabletop guide.
WARNING:
The detector holder must only be mounted such that the
detector holder is positioned above the surface of the
table.
c) Fasten the detector holder by turning the knob thoroughly clockwise.
3.
Adjust the height of the detector holder.
a) Hold the detector holder with one hand and loosen with the other
hand the knob that secures the arm.
254
| DR 800 | Options and accessories
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...