Troubleshooting list
►
Have defective products repaired by Aesculap Technical Service, see Technical Service.
Fault
Cause
Detection
Remedy
Product not running
No battery
No battery in the battery shaft
Insert battery.
Battery not charged
No signal tone when inserting
the battery
Charge battery in charger.
Battery broken
No signal tone when inserting
the battery
Have the manufacturer repair the battery.
Product in secured OFF position
Safety catch is in OFF position
Turn the safety catch into the ON position.
Product defective
Product not running
Have the manufacturer repair the product.
Product becomes too hot
Excessive use
Product heating
Observe instructions for use (nominal operating
mode).
Reprocessing/maintenance carried
out incorrectly
Product heating
Observe instructions for use (processing, care).
Preventive measure: Oil product before every ster-
ilization.
Fall damage, product defective
Product heating
Have the manufacturer repair the product.
Blunt tool
Tool and product become hot
Change tool.
Insufficient power
Product broken
Insufficient product perfor-
mance
Observe instructions for use (processing, care).
Preventive measure: Oil product before every ster-
ilization.
Have the manufacturer repair the product.
Steep rise in temperature after a
short period of time
Observe instructions for use (nominal operating
mode).
Have the manufacturer repair the product
Blunt tool
Tool blades worn
Change tool.
Loud running noise
Broken gearing/ball bearings of the
product
Loud, perceptible noise during
operation
Follow operating instructions (preparation, care).
Preventative: lubricate product before every steril-
ization.
Have the manufacturer repair the product
Lid cannot be
attached/removed
Lid not compatible
Lid does not click into place
Use the compatible lid for GA334.
Lid deformed/broken
Lid is difficult or cannot be
mounted/removed
Have the manufacturer repair the lid.
Connection on product
deformed/broken
Lid is hard or impossible to
install/remove
Have the manufacturer repair the product.
Battery cannot be
inserted/removed
Battery not compatible
Battery cannot be completely
inserted into the battery com-
partment
Use the compatible battery for GA334.
Battery deformed/broken
Battery is hard or impossible to
insert/remove
Have the manufacturer repair the battery.
Battery shaft to product
deformed/defective
Battery is difficult or cannot be
mounted/removed
Have the manufacturer repair the product
Sterile funnel or battery
removal device cannot be
attached
Sterile funnel or battery removal
device not compatible
Sterile funnel or battery removal
device not attachable to the bat-
tery shaft
Use suitable sterile funnel or battery removal
device for GA334.
Sterile funnel or battery removal
device deformed/broken
Sterile funnel or battery removal
device difficult or not attachable
Change sterile funnel or battery device.
Battery compartment at product
deformed/broken
Sterile funnel or battery removal
device is hard or impossible to
attach
Have the manufacturer repair the product.