Manual cleaning with wipe disinfection
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT:
Room temperature
*
Suitable enzyme solution: Helizyme, Cidezyme (the latter used for validation purposes)
►
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
►
Push the saw blade coupling
14
towards the housing as far as it will go.
Phase I
►
Mobilize non-rigid components during cleaning, e.g. pull locking
sleeve
20
back from the support foot
16
.
►
Clean the product under running water, using a suitable plastic clean-
ing brush until all visible residues have been removed from the sur-
faces.
►
Use cleaning template
15
to clean the slot of the saw blade
coupling
14
. Insert cleaning template into the slot and carefully push
out the debris through the slot or the inspection holes at least 5 times
in both directions. Turn template 180° in order to do so.
►
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
►
Use rinsing adapter
21
to clean the inside of the saw blade
coupling
14
:
– Fit rinsing adapter
21
onto rotating sleeve
12
.
►
Using an appropriate syringe, rinse slot manually with 100 ml tap
water.
►
Remove rinsing adapter.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
Phase II
►
Pull back locking sleeve
20
on support foot
16
, push cleaning
template
15
between sleeve and support foot and release locking
sleeve
20
.
►
Follow the operating instructions of the enzyme cleaner with regard to
correct concentration, dilution, temperature and water quality.
►
Spray products with a pH neutral enzyme solution, let soak in for at
least 2 minutes and then wipe off.
►
Contamination should be removed with a lint-free cloth or soft brush
moistened with enzyme cleaner.
►
Flexible components (such as sleeves) should be rinsed for 20 seconds
with the water pistol (cold water, at least 2.5 bar).
►
After manual cleaning, check visible surfaces and areas of flexible com-
ponents for residues.
Phase III
►
Rinse product under running tap water for at least 5 minutes.
►
Flexible components (such as sleeves) will move during cleaning.
►
If necessary, repeat the cleaning process (phase 1 to 3).
►
Remove cleaning template
15
from support foot
16
.
Phase IV
►
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
Phase V
►
Wipe all surfaces of the product with a single-use disinfecting wipe.
Phase VI
►
Rinse disinfected surfaces after the prescribed reaction time for at least
1 minute under running demineralized water.
►
Drain any remaining water fully.
Stage VII
►
Dry the product in the drying phase with suitable equipment (such as
lint-free cloths, pressurized air).
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemical
I
Pre-cleaning
RT (cold)
≥
2
–
D–W
Until visually clean
II
Cleaning with enzyme solution
RT (cold)
≥
2
0.8
D–W
pH neutral*
III
Intermediate rinse
RT
≥
5
–
D–W
–
IV
Drying
RT
–
–
–
–
V
Wipe disinfection
–
>1
–
–
Meliseptol HBV wipes with 50 % propan-1-ol
VI
Final rinse
RT (cold)
0.5
–
FD–W
–
VII
Drying
RT
–
–
–
–
Summary of Contents for Acculan 4
Page 2: ...18 16 19 20 21 11 14 22 2 3 1 13 17 8 9 4 5 12 15 7...
Page 3: ...B C 6 4 1 10 4 1 22 14 A D 7...
Page 117: ...115 ru 6 3 6 3 1 2 1 7 7 1 B Braun eIFU eifu bbraun com Aesculap 7 2 6 45 C...
Page 122: ...120 ru GA334 GA334...
Page 123: ...121 ru GA334 ON 7 Aesculap 180...