6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.5 Measurement Function
5-29
5.2.5.2
Measurement Ranging
The measurement range is determined by the relationship between the measurement auto range ON/OFF
and the source/measurement function.
NOTE:
For the pulse source or pulse sweep source mode, the range is always fixed even when the measurement
auto range is set to ON.
1. Operating range for the measurement auto range
When the measurement auto range is enabled, the upper and lower range levels are within
20 digits
of the values shown in the table below.
Source function
Measurement auto range OFF
Measurement auto range ON
Voltage measurement
Current measurement
Voltage measurement
Current measurement
Voltage source
Fixed to the source range Fixed to the limiter range Fixed to the source range
Current Source
Fixed to the limiter range Fixed to the source range
Fixed to the source range
: Auto range operation enabled (operates with the limiter range as the maximum)
Measurement function
Range
Auto range level
DOWN
UP
Voltage measurement
(Positive)
5 V
-
5.01000
50 V
04.9999
50.1000
250 V
049.999
-
Voltage measurement
(Negative)
5 V
-
1.01000
50 V
00.9999
10.1000
250 V
009.999
-
Current measurement
(Positive/Negative)
3
A
-
3.21000
30
A
02.9999
32.1000
300
A
029.999
321.000
3 mA
0.29999
3.21000
30 mA
02.9999
32.1000
300 mA
029.999
-