5.2.11 Operating Multiple 6247G/6247C
5-60
•
Operational Timing
2. Restrictions on setting
•
The 6247G/6247C has a Tsync (approx. 30
s) time delay from the external trigger input to the
measurement start. Consider this time delay when using two or more 6247G/6247 units in syn-
chronous operation.
•
For all three units, set the measurement range to Fixed and auto zero to OFF.
•
Using the external trigger has restrictions on setting Tp and Th for slaves. (Refer to Section
5.2.10.1).
•
The first synchronous sweep step has a gap within the Th accuracy range.
OPERATE OUT
SYNC OUT
Source and
measurement
Tds1
Td1
Tp1
Tds2
Td2
Tp2
Tds3
Td3
Tp3
Source and
measurement
Source and
measurement
Synchronous delay time
Tsync=Approx. 30
s
No.1
No.2
No.3