6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.4 Source Function
5-22
5.2.4
Source Function
This section describes restrictions on the source function and operations.
5.2.4.1
Source Mode, Source Function and Setting Parameters
The following shows relationships between the source-related setting parameters.
1. For the DC or pulse source modes, the VS and IS parameters can be changed regardless of the func-
tions currently set.
2. For the DC or pulse sweep modes, only function parameters currently set are changeable.
Pulse source
DC source
Pulse source
DC source
Pulse sweep source
DC sweep source
Pulse sweep source
DC sweep source
Linear sweep
Fixed sweep
Random sweep
Voltage random
source value
(8000 data)
Current random
source value
(8000 data)
Current pulse sweep base value
Voltage pulse sweep base value
Voltage stop address
Voltage start address
Voltage sampling count
Voltage level value
Voltage step value
Voltage stop value
Current stop value
Current step value
Current level value
Current sampling count
Current start address
Current stop address
Current start value
Current bias value
Voltage LO limit value (*2)
Voltage HI limit value (*2)
Current HI limit value (*1)
Current LO limit value (*1)
Voltage bias value
Voltage start value
Voltage pulse base value
Current pulse base value
Current source value
Voltage LO limit value (*2)
Voltage HI limit value (*2)
Current HI limit value (*1)
Current LO limit value (*1)
Voltage source value
(*1) (*2): The limit values are shared by the DC, pulse, DC sweep and pulse sweep source modes.
2-slope linear sweep
Current last value
Voltage last value
Current 1st step value
Voltage 1st step value
Current first value
Voltage first value
Current middle value
Voltage middle value
Current 2nd step value
Voltage 2nd step value