5.2.13 Error log
5-66
5.2.13
Error log
The 6247G/6247C holds the error number in the error log memory when it detects an error.
1. Operation
A maximum of five memory areas are available for error logs and they operate as follows:
2. Clearing an error log
The error logs are cleared in the following conditions and the ERR indicator goes OFF. However, the
error logs are not cleared by the *RST command.
•
Power ON
•
When reading the error logs (The error logs are cleared after displaying the error log screen and
completing the menu screen.)
•
When the ERR? or *CLS command is executed.
3. Reading out error logs
Select the parameters,
M) SYSTEM
and
5) Error Log
on the menu screen.
The number of errors is displayed.
Press
to display the error contents.
Turn
to change the error log number.
An error message such as the one below appears if the entered value is set out of range.
• A maximum 5 error numbers are stored in the order of detection.
• If detection exceeds more than 5, then the fifth error log is overwritten by the last error.
• The ERR indicator goes ON when an error log is stored.
5
4
3
2
1
Order of
storing
Overwrites the 6th or later
error here.
Number of error logs
Error log