6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.4 Source Function
5-27
2. Setting the output resistance in Suspend status
Set selecting
A) SOURCE
and then
3) Suspend Z
in the menu.
HiZ:
High resistance output status. The current limit values are set to 100 nA (3
A range).
LoZ: Low resistance output status.
VS: The setting current limit values are applied.
IS: The current limit values are set to 100 digits (100 nA) when the IS range is 3
A
range, or 30 digits when it is other than 3
A range.
3. Shifting between Operate, Standby, and Suspend
NOTE:
The source data during sweep is generated with the following timing.
1. Standby
Operate
2. Standby
Suspend
3. When Operate is set after changing the sweep parameters in Suspend status
Standby status
DC source/Pulse source
Operate status
Suspend status
Power ON
OPR
STBY
OPR
STBY
SHIFT
+
SUSPEND
OPR
Standby status
Sweep source
Operate status
Suspend status
Power ON
OPR
STBY
OPR
STBY
SHIFT
+
SUSPEND
Bias output status
Sweep start
Sweep status
Sweep stop
OPR