6247G/6247C DC Voltage Current Source/Monitor Operation Manual
A-8
1. Measurement execution time
Conditions: Source range; fixed
Measurement range; fixed, Trigger mode; external trigger,
Number of measurement digits; 5½ digits
Integration time; 100
s, Measurement delay; 0.3 ms, Source delay; 30
s
Period; 2 ms, Pulse width; 1 ms
Header; OFF, Block delimiter; EOI (DL2)
•
Time from measurement by trigger input (*TRG) to completion of data output to GPIB/USB
•
Time from measurement by source command reception and trigger input (*TRG) and to com-
pletion of data output to GPIB/USB
In the DC or pulse source mode
•
Time from measurement by spot command reception (measurement trigger after setting the
source value for the setting source function) to completion of data output to GPIB/USB
In the DC or pulse source mode
Source value condition
GPIB
execution
time
USB
execution
time
In DC or pulse source
5 ms
18 ms
When the sweep start value is output
8 ms
19 ms
When the sweep step value is output
6 ms
19 ms
Source
Command
GPIB
execution
time
USB
execution
time
Voltage source
SOV<data> (<data>: 1 character)
8 ms
23 ms
Current source
SOI<data> (No unit, <data>: 3 characters)
8 ms
25 ms
Source
Command
GPIB
execution
time
USB
execution
time
Voltage source
G<data> (<data>: 1 character)
6 ms
19 ms
Current source
G<data> (No unit, <data>: 3 characters)
7 ms
21 ms