6247G/6247C DC Voltage Current Source/Monitor Operation Manual
6.6 Data Output Format (Talker Format)
6-24
2. Mantissa part and exponent part
The exponent column in the list below shows cases of the scaling calculation not being performed.
Measurement function
Unit display
Decimal point and unit
symbol form
Exponent form
Mantissa part
Exponent
part
Mantissa part
Exponent
part
DC voltage
measurement
Measurement
range
5 V
d.dddddd
E+00
d.dddddd
E+00
50 V
dd.ddddd
E+00
E+01
250 V
ddd.dddd
E+00
E+02
DC current
measurement
3
A
d.dddddd
E-06
E-06
30
A
dd.ddddd
E-06
E-05
300
A
ddd.dddd
E-06
E-04
3 mA
d.dddddd
E-03
E-03
30 mA
dd.ddddd
E-03
E-02
300 mA
ddd.dddd
E-03
E-01
Resistance
measurement
Significant
digits
1 digit
0000.0d
E-09 to
E+09
00000d.
E-08 to
E+11
00000.d
d.
2 digits
0000.dd
0000d.d
d.d
dd.
3 digits
d.dd
000d.dd
dd.d
ddd.
4 digits
d.ddd
00d.ddd
dd.dd
ddd.d
5 digits
d.dddd
0d.dddd
dd.ddd
ddd.dd
High limit value detected in resistance measurement *1 +9.99999
E+37
+9.99999
E+37
Low limit value detected in resistance measurement *1 +9.99999
E+36
+9.99999
E+36
Range over
9.99999
E+35
9.99999
E+35
IS is below 20 digits or IM is below 200 digits. *1
+9.99999
E+34
+9.99999
E+34