6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.5 Measurement Function
5-33
5.2.5.3
Measurement Delay Time and Measurement Value
1. Measurement at the pulse value timing
When displaying the resistance value, the calculation is made by the measurement value and the pulse
value.
2. Measurement at the base value timing
When displaying the resistance value, the calculation is made by the measurement value and the base
value.
3. Measuring at the timing overlapping with the pulse value and the base value
The measurement value changes depending on the time ratio of the pulse value to the base value in
the measured time. The resistance value is calculated from the measurement value and pulse value,
which makes the value inaccurate.
Pulse measurement
Tp
Base value
Pulse value
Base value
Td
Measurement
Tp
Base value
Pulse value
Base value
Measurement
Tp
Base value
Pulse value
Base value
Measurement