6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.8 Source and Measurement Timing
5-42
•
Source delay time (Tds) + pulse width (Tw) + 300
s < period time (Tp)
•
Period time (Tp)
600 ms
Conditions: Source mode
Pulse mode or pulse sweep mode
Integration time Sample hold mode
2. Relationship between restrictions and source mode
The table below shows the relationship between the restrictions described above and the source mode.
3. Source delay time, measurement delay time and pulse width
The setting resolutions of the source delay time, the measurement delay time and the pulse width are
determined by the period time resolution. Values rounded off to the resolution are set.
NOTE:
Even if the measurement is set to OFF, the restrictions on the measurement delay time are valid.
Restriction
Source mode
DC
PLS
DC-SWP
PLS-SWP
[Tds + 300
s < Tp]
-
[Td + 300
s < Tp]
[Tds
Td]
-
[Tds + Tw + 300
s < Tp]
-
[600 ms
Tp]
-
-
Period time setting range
Resolution
0.500 ms to 60.000 ms
1
s
60.01 ms to 600.00 ms
10
s
600.1 ms to 6000.0 ms
100
s
6001 ms to 60000 ms
1 ms
Tp
Tds
300
s or more
Tw