6247G/6247C DC Voltage Current Source/Monitor Operation Manual
6.6 Data Output Format (Talker Format)
6-25
3. String delimiter
A string delimiter ", " (comma) is output to specify a boundary in a data stream.
It is used when multiple data sets are read out from the measurement data memory.
4. Block delimiter
A block delimiter is output to specify the end of a unit of data.
The type of block delimiters can be specified by the following commands:
VS is set to 0 (zero). *1
+9.99999
E+33
+9.99999
E+33
Scaling error
9.99999
E+32
9.99999
E+32
TOTAL error
9.99999
E+31
9.99999
E+31
No data exists at recall. *2
+8.88888
E+30
+8.88888
E+30
*1:
This may occur in resistance measurement.
*2:
Data does not exist when the measurement buffer memory data is read out.
Block delimiter
Command
Default
CR LF+EOI
DL0
LF
DL1
EOI
DL2
LF+EOI
DL3
(EOI is a GPIB function. It is not available on USB or RS-232.
As for RS-232, the block delimiter is always CR LF regardless
of the setting.)
Measurement function
Unit display
Decimal point and unit
symbol form
Exponent form
Mantissa part
Exponent
part
Mantissa part
Exponent
part