6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.11 Operating Multiple 6247G/6247C
5-59
5.2.11
Operating Multiple 6247G/6247C
This section describes synchronous operation, serial connection, and parallel connection using more than
one 6247G/6247C.
5.2.11.1
Synchronous Operation
The synchronous operation of multiple 6247G/6247C units allows synchronization of measurement timing
in the DC source mode, as well as allowing synchronization of both source and measurement in the pulse
source mode and the sweep source mode.
The timing control for the synchronization is performed by the external control signals; TRIGGER IN,
SYNC OUT and OPERATE OUT, and the time parameter settings such as measurement delay and source
delay.
1. Three unit synchronous operation using SYNC OUT
•
Setting
•
Connection
Parameter
No.1
No.2
No.3
SYNC OUT control signal
SYNC OUT
-
-
OPERATE IN/OUT control signal
OPERATE OUT
OPR/SUS IN
OPR/SUS IN
Trigger mode
AUTO
HOLD
HOLD
6247G/6247C
No.1
SYNC OUT
OPERATE IN/OUT
TRIGGER IN
OPERATE IN/OUT
6247G/6247C
No.2
TRIGGER IN
OPERATE IN/OUT
6247G/6247C
No.3
(Slave)
(Master)
(Slave)