6247G/6247C DC Voltage Current Source/Monitor Operation Manual
5.2.1 Operations in DC Source Mode
5-9
a.
When the trigger mode is AUTO:
•
The measurement repeats itself at intervals specified by the period time.
•
If the measurement does not finish in the specified period, the period time is extended and the
TpALM indicator goes ON.
b. When the trigger mode is HOLD:
•
The measurement starts when the measurement delay time has passed after trigger input.
•
Trigger input during measurement is ignored.
c.
In Standby or Suspended status:
•
The measurement is not executed in Standby or Suspend status.
Changing the
source value
AUTO
Source value
changing induces
range changing.
Tp: Period time
Td: Measurement
delay time
Tm: Measurement time
(Integration time +
Measurement data
processing time)
Tcn: Operation processing time
Trc: Range change processing
time
HOLD
Table 5-2 Operations in DC Source Mode (2 of 2)
Operational
condition
Trigger mode
Description
Operation
Remarks
COMPLETE OUT [END]
COMPLETE OUT [HI/GO/LO]
Td
Tp
Tp
Td
Tm
Trc
Td
Tm
Previous value
Changes the
source value
COMPLETE OUT [FRONT]
Td
Tm
Measurement
trigger
Changes the
source value
Previous value
COMPLETE OUT [END]
COMPLETE OUT [HI/GO/LO]
TRIGGER IN
COMPLETE OUT [FRONT]
Trc
Tp