Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
9 Troubleshooting
88
COI/AZ30E–EN Rev. D
––S054.011––
CELL WARMING UP
Wait for cell
temp. to reach
690°C/1274°F
Cell temperature is below 690 °C
(1274 °F).
Wait for cell temperature to reach
690 °C (1274 °F) .
––S052.012––
CELL STABILIZING
Wait for cell
temperature to
stabilize (5min)
Cell temperature above 690 °C
(1274 °F) but not stable.
Wait for cell temperature to stabilize
(5 minutes).
––S050.004––
O2 SENSOR
OUT OF RANGE
Address process
issue or extend
operating range
Process oxygen concentration
beyond set operating range.
Address process issue or extend
oxygen operating range.
––S048.005––
TEMP. SENSOR
OUT OF RANGE
Check operating
conditions
Maximum temperature range for
thermocouple exceeded.
Check operating conditions.
Status
Icon
Diagnostic Message
Displayed
Possible Cause(s)
Corrective Measure(s)
Table 9.3 Diagnostics – Out of Specification Messages (Continued)