Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
8 HART®-Protocol
78
COI/AZ30E–EN Rev. D
45
Trim Primary Variable
Current DAC Zero
The transmitter re-calibrates the current output
Zero Offset
based on the externally measured primary value received.
46
Trim Primary Variable
Current DAC Gain
The transmitter re-calibrates the current output
Span Scale
Factor
based on the externally measured primary value
received.
48
Read Additional
Transmitter Status
The status of the transmitter, determined from results of the
continuous self-diagnostics, is reported every time
communications with the transmitter are established.
If the transmitter indicates there is additional status
information, it can be obtained via this command.
49
Write Primary Variable
Sensor Serial Number
Enables a transducer serial number to be entered.
60
Read Analog Channel
and Percent of Range
Returns the current mA value on
Current Output 2
and that
value as a percentage of its range.
66
Enter / Exit Fixed Analog
Channel Mode
Current Output 2
can be set to a constant mA value.
Sending a value of '0x7F, 0xA0, 0x00, 0x00' cancels the
fixed output mode.
67
Trim Analog Channel
Zero
The transmitter re-calibrates the
Current Output 2
/ Zero
Offset
based on the externally-measured primary value
received.
68
Trim Analog Channel
Gain
The transmitter re-calibrates the
Current Output 2
/
Span
Scale Factor
based on the externally-measured primary
value received.
Command
No.
Description
Details
Table 8.3 Common Practice Command Set (Continued)