Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
8 HART®-Protocol
80
COI/AZ30E–EN Rev. D
8.5.2 Field Device Status Second Byte
Second Byte
Description
Bit #7
Field Device Malfunction
A hardware error or failure has been detected by the device.
Further information may be available through the
Read Additional Transmitter
Status
command.
Bit #6
Configuration Changed
A write command has been executed.
A set command has been executed.
Bit #5
Cold Start
Bit #4
More Status Available
More Status information is available than can be returned in the
Field Device
Status
. Command #48
Read Additional Status Information
provides this
additional status information.
Set if any of the device-specific status bits in command #48 are set.
Bit #3
Primary Variable Analog Output Fixed
The analog and digital analog outputs for the
Primary Variable
are held at the
requested value. They do not respond to the applied process.
Primary analog output set to a fixed value via HART (or via the HMI) putting the
current output into test mode.
Bit #2
Primary Variable Analog Output Saturated
The analog and digital outputs for the 'Primary Variable' are beyond their limits
and no longer represent the true applied process.
Calculated
Current Output 1
value is beyond the physical limits of the output.
Bit #1
Non Primary Variable Out of Limits
The temperature or cold junction cell voltage are beyond the operating limits of
the probe. The
Read Additional Transmitter Status
command, #48, is required to
identify the variable.
Set if SV, TV or QV sensor limit failure diagnostic bits are set.
Bit #0
Primary Variable Out of Limits
The measured oxygen value is beyond the operating limits of the probe.
PV sensor out-of-limits diagnostic set.
Table 8.5 Field Device Status Second Byte