Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
6 Programming
58
COI/AZ30E–EN Rev. D
6.4.7 Input/Output
Used to assign relays (1 and 2), digital I/O (1 and 2) and
current outputs (1 and 2).
Menus displayed at this level are dependent on the system
configuration / options fitted.
Parameter
Comment / [Range]
Default
Relay 1 (2)
Relay Assignment
Selects the alarm and diagnostic states required to
activate the relay.
Each
Relay Assignment
parameter can be
Assigned
or
Not Assigned
.
Note
.
Assigned
parameters are OR'd together.
Not
Assigned
(all
parameters)
Alarm 1 (2, 3, 4)
See Section 6.4.6, page 57 for alarm types and
triggers.
Cal. In Progress
Relay is activated when a calibration is in progress.
Cal. Failed
Relay is activated if a calibration fails.
Gas 1 Not Present
Relay is activated if Test Gas 1 is not detected.
Gas 2 Not Present
Relay is activated if Test Gas 2 is not detected.
Test Gas 1 Valve
Relay is activated if Test Gas 1 valve is open.
Test Gas 2 Valve
Relay is activated if Test Gas 2 valve is open.
Diag. – Failure
Activates the relay if a
Failure
diagnostic status is
generated – see Section 9.2.1, page 84.
Diag. – Off Spec.
Activates the relay if an
Out of Specification
diagnostic status is generated – see Section 9.2.2,
page 87.
Diag. – Maint. Reqd.
Activates the relay if a
Maintenance Required
diagnostic status is generated – see Section 9.2.3,
page 89.
Diag. – Chk.
Function
Activates the relay if a
Check Function
diagnostic
status is generated – see Section 9.2.4, page 91.
Table 6.12 Input/Output Menu
7
Menu
Exit
Select
Input/Output